REQUEST A QUOTE or ASK A QUESTION ABOUT THIS PRODUCT
Available Options
  • Code C - Enhanced Performance (±0.02%)
  • Code J - Ultra Precision Performance (±0.01%)

Precision Torque Transfer Standard

Model #RTM 2300DV

  • Ultra-Precise Torque Transfer
  • Accurate Field Calibration of Production Torque Transducers
  • 0.01% Combined Nonlinearity & Hysteresis
  • 0.0003%/deg F Temperature Compensation
  • ±10.000 Vdc Analog Output
  • Digital Output of Real Time, Max/Min and Spread Data
  • PC Interface Software Included
  • No Manual Adjustments

General Specifications

  Code N Performance Code C Performance Code J Performance
Torque Range Factory Set @ Transducer Full Scale Torque.
Units of Measure lbf-in, lbf-ft, ozf-in, ozf-ft, N-m, kN-m, N-cm, kgf-m, kgf-cm, gf-cm
Combined Nonlinearity & Hysteresis (Best Fit Line Basis - See Tech Memo 230104)
≤±0.04% ≤±0.02% ≤±0.01%
Overrange (% of Range) 150, Applies to All Outputs
Overload (% of Range) 200
Repeatability (% of Range) ≤±0.015 ≤±0.01 ≤±0.01
Accuracy Class 0.04 0.036 0.018
Zero Drift (% of Range per oF/per oC) ≤±0.001/0.0018 ≤±0.0003/0.00054 ≤±0.0003/0.00054
Span Drift (% of Reading per oF/per oC) ≤±0.002/0.0036 ≤±0.002/0.0036 ≤±0.001/0.0018
48 Hour Drift (% of Range) ≤±0.03 ≤±0.02 ≤±0.02
Analog Output Signal, Auto Scaled ±10 Vdc with ±15 Vdc overrange or ±5 Vdc with ±15 Vdc overrange
Signal Cutoff Frequency 0.1 Hz to 500 Hz in thirteen 1-2-5 steps plus 3 kHz.
frequency Modulated Output 10 ±5 kHz or 20 ± 10 kHz or 40 ± 20 kHz
Torque Sampling Rate and Bandwidth Sampled at 20 kHz.  Torque 3 db bandwidth is 3 kHz reducable by filters.
RS232/RS422/RS485/USB Communications Com port outputs Torque and Temperature with units of measure. Inputs instrument settings.
Interface Software Provided for PC interface.  Includes 20 foot interconnect cable.
Rotor Temperature
Supply Voltage 10 to 26 Vdc

Please contact factory for latest specifications.