S. Himmelstein and Company  
Site Search:
The World's Best Torque Instruments Since 1960
Search by Torque Rating:

Contact Himmelstein

Model 701 & 711 Strain Gage Conditioners

Model 701 & 711 Strain Gage Conditioners

These instruments are full featured strain gage conditioner/readouts; the Model 701 handles one and the Model 711 handles two transducers. Both provide fast, accurate data for each channel. What's more, 18 of the most useful processing functions and 26 real-time digital calculations are built in. Their use converts the Model 701/711 into powerful production and performance test analyzers with easily configured characteristics. A two line alphanumeric display publishes measured and computed data, units of measure and test status. Transducer calibration is automatic, requiring no manual adjustments.

General Specifications

Transducer(s)

 

Type: Any 80 to 2000Ω strain gage transducer, directly wired or transformer coupled.
Connections:  Provision for 4, 6, or 7 wire circuits.

Maximum Cable Length

500ft for transducer impedance >100Ω and 200ft for transducer impedance <100Ω.

Transducer Excitation

3Vrms, 3030Hz ± 0.01% sine wave. Regulated, and short circuit protected.

Input

Sensitivity: 0.5 to 5mV/V with 50% overrange; automatically scaled.
Impedance: 100MΩ in parallel with 33pF.

Automatic Null Range

In-Phase Signals: ±10% of Full Scale (with 50% overrange), ±60% of Full Scale (with 0% overrange).
Quadrature Signals: ±1mV/V.

Auto Calibration

Dual polarity shunt calibration with provision for CAL resistor feedback.

Spurious Signal Rejection

60Hz: 120dB common mode, 100dB normal mode. Carrier quadrature: 60dB.

Antialias Filter

200Hz, 7 pole Bessel response filter.

Low Pass Filtering

4 pole Bessel response digital filters with 11 cutoff frequencies from 0.1 to 200Hz in 1-2-5 steps.

Signal-to-Noise Ratio1

 

@ 1mV/V F.S.: 86/76/66/62dB with 1/10/100/200Hz filters.
@ 5mV/V F.S.: 86/80/72/66dB with 1/10/100/200Hz filters.

Resolution

0.01% of Full Scale.

Overall Accuracy

(at 77°F/25°C) 0.02% of Full Scale, worst case.

Temperature Effects

Zero: ±0.001% of Full Scale/°F (max); Span: ±0.001% of Full Scale/°F (max).

Display

Type: 2 line by 16 alphanumeric characters, each 0.2" wide by 0.3" high. Backlit LCD with adjustable contrast.
Views: Select either 2 Channels, 1 Channel with Limit Status, or 1 Channel with I/O Status.
Data Displayed: Select from Current, Max, Min, Spread, Held data and Tare value.
Data Format: Engineering units with 6 digits (1-2-5 format) and 5 character user-entered legend/descriptor.

Number of Channels

Hardware: Supports one (Model 701) or two input channels (Model 711).
Calculated: One (CH3). Choose from 26 formulas based on CH1, CH2, and a constant.

Response (per channel) Data Sampling & Max/Min Update Rates: 2000Hz (hardware channels), 50Hz (CH3 calculation).
Limit Checking Rate: 1000Hz (hardware channels), 50Hz (CH3 calculation).
Logic I/O Response Time: 1ms (hardware channels), 20ms (CH3 calculation).
Update Rate for each Analog Output: 1000Hz.
Control

(All I/O functions can be OR'd in any combination. The pattern function adds ANDing capabilities.)

Input Actions/channel: Logic inputs, outputs, and internal Matrix signals control following actions. Tare, Clear Tare, Hold, Clear Hold, Reset Max/Min, Clear Latched Limits, Check Limits, Do Max/Mins, Apply +CAL, Apply -CAL.
Output Events/channel: The following events drive Logic outputs and internal Matrix signals. HI Limit, NOT HI Limit, IN Limit, NOT IN Limit, LO Limit, NOT LO Limit, At Max, NOT At Max, At Min, NOT At Min.
Three User-defined Patterns: Patterns drive Logic outputs and internal Matrix signals.
Patterns are based on Logic inputs, outputs, and internal Matrix signals.

Limit Checking

Each channel has a HI and LO limit which may be latched or unlatched, absolute or signed, and with or without hysteresis. Select either Current, Max, Min, Spread or Held data for limit checking.
Limit violations on any or all channels can be set to trigger backlight flashing in any of the display view modes.

Four Logic Inputs

(each with programmable function -- destination)

Type: TTL compatible, Schmitt Trigger, low-true with 47kΩ pull-up. Input current is -100µA @ 0V.
Protection: to "±130VDC (or 130VAC).

Six Logic Outputs

(each with programmable function -- source)

Type: Open collector, low-true. Operating @ 24V (max) and 0.3A max sink current.
Protection: Short circuit (current and thermal limits) and overvoltage (0.5A fuse).

Serial Communication Port

(user selectable as RS232, RS422, or RS485)

BAUD Rate: 300 to 38400. Maximum Cable Length: 4000ft (RS422/RS485), 50ft (RS232).
Maximum Number of Devices: 32 (RS485), 1 (RS232/422).
120Ω Termination Resistors (RS485): user selectable for RXD and TXD.
RS422/485 Transceivers: Slew-rate limited, short circuit protected (current & thermal limits).
RS232 Drivers: Short circuit protected (current limit).
Serial I/O’s: Use a 9 pin D connector. They are ±15kV ESD protected and float (100kΩ) with respect to Earth Ground.
Commands: Control of all modes, settings, and measurements.

Non-Volatile Memory Storage for System Settings

EEPROM, batteries are not used.

External +5VDC Power

(on I/O connector) 250mA, short circuit (current limit) and overvoltage (1A fuse) protected.

Dual Analog Outputs

(each assignable to any of the displayed channels)

Output Impedance/Minimum Load Resistance: <1Ω/10kΩ.
Full Scale: ±5V or ±10V (user selectable). Resolution is ±2mV @ ±5V FS or ±4mV @ ±10V FS.
Overrange: ±8.2V @ ±5V FS or ±13.5V @ ±10V FS.
Non-linearity: ±2mV @ ±5V FS or ±4mV @ ±10V FS.
Overall Error (worst case, including temperature effects): ±5mV @ ±5V FS or ±10mV @ ±10V FS.
Filter: 100Hz, 5 pole Bessel response low pass filter.
Protection: Short circuit (current limit) and overvoltage (0.25A fuse) protected.

Size and Weight 6.5" wide, 2.9" high, 8.7" deep. Weight is 3 pounds.
Operating Temperature

+41°F to +122°F (+5°C to +50°C).

Input Power

90VAC to 250VAC, 50/60Hz @ 25VA, max. Two 2A/250V fuses, line filter, and rear power switch.

Notes:

1. The ratio expressed in decibels (dB), of Full Scale to noise spread. Measurements are made for a 1 minute interval using a 350Ω bridge.

2. Specification is subject to change without notice.

 